Abstract
Twiddler syndrome is a rare complication of cardiac implantable electronic devices (CIEDs) typically caused by patient manipulation of the generator, leading to lead dislodgement. We present a unique case of a patient with a biventricular implantable cardioverter-defibrillator (ICD) who developed atrial and ventricular lead dislodgement without evidence of device manipulation. The patient experienced six consecutive inappropriate ICD shocks due to oversensing and double counting of atrial and ventricular signals by the right ventricular (RV) lead. This case highlights a variant of Twiddler syndrome occurring in the absence of twiddling, underscoring the importance of vigilance for device malfunction even without patient-induced factors.Keywords:
Twiddler syndrome Implantable cardioverter-defibrillator (ICD) Lead dislodgement Device malfunction Inappropriate shocks Cardiac implantable electronic device (CIED) Lead migration
This work is licensed under a Creative Commons Attribution 4.0 International License. To view a copy of this license, visit
http://creativecommons.org/licenses/by/4.0/
Figures
References:
| [1] | Bayliss CE, Beanlands DS, Baird RJ. The pacemaker-twiddler’s syndrome: a new complication of implantable transvenous pacemakers. Can Med Assoc J. 1968; 99(8): 371–373. |
| |
| [2] | Ellenbogen KA, Wood MA, Shepard RK. Delayed complications following pacemaker implantation. Pacing Clin Electrophysiol. 2002; 25(8): 1155–1158. |
| |
| [3] | Nicholson WJ, Tuohy KA, Tilkemeier P. Twiddler’s syndrome. N Engl J Med. 2003; 348(17): 1726–1727. |
| |
| [4] | Li A, Kaura A, Sunderland N, Dhillon PS, Scott PA. The Significance of Shocks in Implantable Cardioverter Defibrillator Recipients. Arrhythm Electrophysiol Rev. 2016 Aug; 5(2): 110-6. |
| |
| [5] | van Rees, J, Borleffs, C, de Bie, M. et al. Inappropriate Implantable Cardioverter-Defibrillator Shocks: Incidence, Predictors, and Impact on Mortality. JACC. 2011 Feb, 57(5) 556–562. |
| |