Copyright © 2020 Science and Education PublishingCite this paper: Myron Chernets. Calculation Method for Assessing Contact Parameters in the Hip Prosthesis Made of Thermo-diffusion Nitrided Grade 2 / ultra-high Molecular Weight Polyethylene.
Journal of Biomedical Engineering and Technology. 2020; 8(1):6-13. doi: 10.12691/jbet-8-1-2.
Correspondence to: Myron Chernets, Lublin University of Technology, Faculty of Mechanical Engineering, 20-618 Lublin, Poland. Email:
myron.czerniec@gmail.com, m.czerniec@polluAbstract
In this paper, the author proposes a new calculation method for calculating contact parameters (i.e., maximum contact pressures, angle and diameter of contact) in the hip prosthesis made of thermo-diffusion nitrided (TDN) Grade 2 and ultra-high molecular weight polyethylene (UHMWPE). The paper investigates the impact of hip joint load, prosthesis head diameter and radial clearance on the above contact parameters. Relationships between maximum contact pressures and the above-mentioned contact parameters are determined. Both increasing radial clearance and endoprosthesis loading cause a linear increase in contact pressure. However, when the head diameter increases, there is a non-linear reduction of contact pressure. The contact diameter increases linearly as the head diameter increases. According to the given method, the endoprosthesis with non-spherical surfaces of its elements was also tested (Alpharabola geometry). An analysis of the effect of head deviation from sphericity in the form of oval on contact pressure, contact angle and contact diameter was performed. Beneficial effects of this geometry were determined.
Keywords